TAKChill - Economical Heat Exchange For Analytical Instruments

  1. Features

    takchill
    1. Utilizes the coaxial tube-in-tube concept to provide the maximum amount of exposed surface area in the smallest possible space.
    2. Unique design incorporates a coiled 1/8" thin wall Teflon line inside a ½" coiling loop.
    3. Chemical line is a continuous tube with no connections which eliminates the possibility of leaks outside the users contained areas.
    4. Unit comes with six feet of additional chemical line on both the input and output sides to facilitate connections from the chemical source to the analytical instrument input.

  1. Meter Info

    takchill
    1. Includes a TC meter to measure outgoing temperature of the chemical line.
    2. Meter provides relay contacts corresponding to the programmed alarm limit.
    3. Standard TC wire included allows for remote mounting of the meter up to six feet away.

Specifications

Water Temp Chemical Inlet Temp Chemical Outlet Temp
70° 57°C 40°C
40°C 27°C
60° 50°C 30°C
40°C 26°C
  1. Additional Info

    1. Chilled Water: 1/2" Teflon tubing for both inlet and outlet; minimum 70°F at max 5LPM
    2. Cooling Rates: Chemical flow @ 35 ml/m (data reflects single loop configuration)
    3. Meter:
      1. Power - Fused (420ma/250V) 24vdc, 3watts
      2. NOTE: External fuse required. (see meter documentation)
    4. Dimensions:
      1. Heat Exchanger - 13"x6"x7"
      2. Meter - 3.3"x1.9"x1.9"